CrystalControl Operation Manual
Modules: Current, Voltage, & Switch Test (MA20)
Note: The MA20 module requires CrystalControl 1.2 or greater, and Chassis  rmware R080005 or greater to operate properly. The MA20 Module Instructions
(P/N 4339) provide further details if required.
The following features are unique to the Current, Voltage, & Switch Test Modules (MA20):
mA: In addition to the Userspan, this module allows for an o set value to be entered
VDC: In addition to the Userspan, this module allows for an o set value to be entered.
!
CAUTION: Changing the Userspan or o set value changes the calibration of the module. Changes in userspan or o set values should only be made by
quali ed personnel and equipment.
Setup Wizard: Select this button to Setup the Switch Test mode. The wizard will test the open and closed switch states to ensure accurate readings in the
eld.
Note: You may use the included nVision Test Leads during the setup process.
Typical view of MA20 Module features in CrystalControl.