Overview 1
Type T Pump Operation Manual
Overview
INTRODUCTION
The Type T Series hand pump, mounted on a drip pan with a manifold assembly, is capable of producing pressures up to 1000 bar/15 000 psi.
These units are self-contained pressure sources designed to test instruments. They may be used to calibrate pressure gauges or to set hydraulic relief valves
and pressure switches. These units produce pressure to test components or systems, and work independently of other pressure sources.
The Type T Series hand pump may be ordered as part of a Pump System, complete with a JOFRA or Crystal Pressure Indicator. Type T Series Pump Systems
include the most commonly used pressure ttings, seals, etc. All packaged in a carrying case with custom insert.
X
Type T Series Pumps
Sealing Compound Medium (Shipped Empty) Pump System
T-1 Buna-N Water (50/50 mixture of Water and Isopropal Alchohol) FWB
T-1-CPF Viton® Oil or Water (Empty) FOV or FWV
T-1/VITON Viton® Water (50/50 mixture of Water and Isopropal Alchohol) FWV
T-1/EPT EPT Water (50/50 mixture of Water and Isopropal Alchohol) FWE
T-1/OIL Buna-N Oil (Oil) FOB
T-1/OIL/VITON Viton® Oil (Oil) FOV
T-1/OIL/EPT EPT Oil (Oil) FOE