FastCalXP Operation Manual
Save Certi cates To
The calibration certi cates created by FastCalXP will be saved in the directory speci ed in Save certi cates to. This directory is set the  rst time the software is
run after installation or when a new database is created and can be changed here using the Browse button.
Certi cate Naming Template
The Certi cate naming template is used by FastCalXP to name the calibration certi cates. In general, certi cates are numbered with the Next certi cate
number incrementing automatically each time a certi cat
e is created. Use number signs (#’s) in the Certi cate naming template to de ne the number of digits
to be used.
Calibration Representative and Quality Representative
The Calibration representative and the Quality representative must be listed users and can be selected using the pull-down lists. In the certi cate template
provided with the software these names will be listed, with signature lines, on the  rst page of each certi cate.
Calibration Settings
The Calibration Settings de ne the default Calibration Method.
Actual
Uses the pressure reading from the reference at the moment the Next Point button is clicked.
Peak
Automatically advances to the Next Point when a peak is detected.
Head Correction
If the gauge and the reference do not sit at the same height, a Head correction can be applied. Enter a correction to be added to the reference pressure, or use
the Correction calculator to calculate a correction given the speci c conditions of the calibration. Refer to the discussion of the Head Correction Calculator
on page 32.