Managing the Database 37
FastCalXP Operation Manual
Use Custom Certi cate Template
When Use custom certi cate template is checked, you can override the default Excel template de ned in the Options dialog. This allows di erent manufactur-
ers or models to have unique certi cate formats.
Cal Interval
Enter the Cal interval (in days) which will be used for calibration scheduling.
Transmitter
Select Transmitter when you will select which parameters of the transmitter to measure during calibration.
Measure mA
Select Measure mA when calibrating the analog output of the transmitter. Specify the Low/High pressure range as well as the corresponding Low/High mA
output of the transmitter.
Measure HART
If the model is a smart transmitter, select Measure HART to view digital output of the transmitter. If the transmitter has an external display check Measure
Displayed to include the external display in the certi cat
e. You can select either Measure Displayed or Measure HART but not both for a model.
Accuracy
There are three ways to de ne the accuracy of the model. Many gauges will have an ASME or ISO/IEC grade which can be selected from the pull-down list. If
the accuracy is not listed, you may be able to simply enter a % of span accuracy.
Alternativ
ely, select Custom to de ne a custom accuracy by specifying a series of pressure ranges and ac
curacies associated with those ranges. Click the Add
button to include a range and accuracy in the speci cation.
When the accuracy is fully speci ed, click OK.