Managing the Database 39
FastCalXP Operation Manual
TEST PROFILES
The Pro les editor provides access to all test pro les stored in the database.
The Pro le Details box is also available for creating a new pro le when a new model is created using the New Gauge Wizard.
Name
When creating a new pro le, it will help to make the Name descriptive and speci c, particularly if you will be using several similar pro les for di erent models.
Units of Measure
Use the pull-down list
to specify the correct Units of measure.
Number of Excercise Cycles
Enter a non-zero Number of exercise cycles to remind the operator to exercise the gauges before beginning any calibration that uses this pro le.
To Create a Uniform Series of Pressure Points
1 Click on Ascending/descending points.
2 Specify the minimum and maximum range of the pro le and the number of asc
ending and descending points.
The Pressure list will be  lled automatically.